Product

Maximum Service Temperature Apparatus - Digitech Roorkee

Maximum Service Temperature Apparatus

Maximum Service Temperature Apparatus (DG-IS-MSTA), is a microprocessor based instrument designed to measure and display the Thermal Resistance of loose filled insulating materials, using hot plate method as per guidelines of Standard: EN 14706. The instrument is automatic and programmable; designed with latest integrated circuit technology. The instrument displays the measured value directly in engineering units.

The samples of the material to be tested are placed on top of a horizontal hot plate fitted on a furnace. The furnace consists of a heater and a sample loading and thickness measuring arrangement. The building blocks of the thermal resistance apparatus is a data logger cum temperature controller. The temperature measurement is done by embedded thermocouples on a hot plate. Five thermocouples are mounted on a hot plate and one thermocouple is provided to monitor the temperature of other locations, as desired.  All channels are displayed on the local display. The voltage & current fed to the heater are also displayed on analogue meters.

Real-time, working temperature is programmable. The Temperature Controller controls the temperature of the hot plate. The temperature setting may be done continuously throughout the operating range. The instrument is mains operated and supplied completely ready to operate along with Operating and Instructions Manual. Suitable window based software is provided to collect data from data logger and provide R-value with the trend graph and report in PDF format.

Window based software:

Window based software is provided to program / monitor and calculation of K-value. Online data and the graph is displayed for all the channels along with time stamp.

Specifications

Product Code DG-IS-MSTA
Category Import Substitute
Heating Area 300 X 300 mm
Overall size 630 X 60 X 260 mm
Plate Thickness >= 10 mm
Material  SS-310
No of thermocouples Total 5 thermocouples are mounted on top side of hot plate. One thermocouple is provided for temperature monitoring of sample.
Thickness deviation measurement By LVDT 0- 10.00 mm
Working Temperature 200 to 1000 °C
Maximum working temperature 1000°C

Data Logger cum Temperature Controller

Display  LCD display
Key Pad 20 keys.
Sensor K Type thermocouples.
Measurement Range 0°C to 1200°C
Resolution 10°C
Measurement accuracy ± 0.5% or better.
Temperature control accuracy ± 2% or better.
Control setting Guarded Hot Plates 200 to 1000 °C
Temperature Setting  KEYBOARD
Sample thickness 25 – 100 mm
Heating rate 50 or 300°C /Hrs.
Computer Interface Serial port RS 232C, transfer rate 9600
Power 230V ac mains 5KW.
Enquiry

Enquiry